Microcuvette ad Hemoglobin Analyzer
Microcuvette pro Hemoglobin analyser
Hemoglobin analyser microcuvette
Product Details:
Material: polystyrene
Vita Shelf: 2 years
◆Storage temperatus: 2°C35°C
Relativum Umor≤85%
Weight: 0.5g
Packing: pieces L / utrem
Valorem positivum / range Reference Range Reference:
◆Adulti masculini: 130-175g/dL
◆ Adultae feminae: 115-150g/dL
◆Infant: 110-120g/dL
Puer: 120-140g/dL
Test Proventus
Mensura ostentationis amplitudo est 0-250g/L.Coagulatio potest specimen sanguinis facere ut microcuvettam impleat, inde in mensuras falsas.
Hemolysis test eventus afficere potest.
Modus Test limitatio
◆ Diagnosis et curatio non solum in experimento eventi confidere debent.Historia clinica et aliae probationes laboratoriae considerari debent
Euismod Specification
Blank.≤1g/L
Repeatability.intra range 30g / L ad 100g / L, SD *≤3g/L;intra range 101g/L ad 250g/L, CV≤1.5%
linearity.intra range 30g/L ad 250g/L, r*≥0.99
Accuracy.Correlatio coëfficientis (r) comparationis experimentum est≥0,99, et relativum declinationis est≤5%
Inter-batch difference≤5g/L
Test procedendi EDTA sanguinem Testis:
- Exempla reposita reddenda sunt ad cella temperies et permixta ante probationem.
◆ Micropipetto vel pipetto utere ut non minus quam 10μL sanguinis hauriatur in vitreo mundo vel alia superficie munda hydrophobica.
Apice reagens ad exemplum contactum, specimen sub actione capillari intrat et partem gerentis implet.
- Sollicite absterget omnem excessum specimen in superficie microcuvette.
◆ Pone microcuvettam in microcuvette possessoris hemoglobini analystoris et tunc possessorem in analysentem ad mensurae initium impellere.